MINIDAR 2

PORTABLE MINI-VENTILATOR "MINIDAR 02-"Avrora"

Designed for controlled artificial lung oxygen ventilation of child (weight from 10 to 30 kg) with active breath, passive exhalation and additional gas stream possibility at a place of an accident, at home, during transportation, as well as at medical institutions.


Composition of the apparatus


ventilation unit;
power supply unit (2 liter cylinder);
supply tube;
set of instruments and accessories;
bag.

Technical parameters

No.

Denomination

Value

1

Adjustment range of a minute ventilation during artificial lung ventilation with oxygen, l/min


0,1 - 10

2

Adjustment range of additional flow (manual APV) of oxygen, l/min

0,1 - 20

3

Adjustment range of ventilation frequency, 1/min

10 - 35

4

Inspiration-to-expiration duration ratio

1:1,5-1:2

5

Controllable output strength, kPa

0 - 1.5

6

Maximum working (safe) gas pressure, kPa

5.0

7

Power supply of the apparatus

Power supply unit, transport cylinder, pneumatic system (0,4+0,05 MPa)

8 Continuous working time if oxygen supplied 24 hrs

9

Overall dimensions, mm:
ventilation unit;
pneumatic power supply unit.


119 x 82 x 115
416 x 212

10

Mass is not more than, kg:
ventilation unit;
complete set.


0.45
7.0

MINIDAR-01-Avrora
XEIN-Avrora
PNEUMOCOMP-04
AINp-01/10-Avrora
Sales


   
control systems for surface ships, control systems for submarines, combat information and control systems, simulators, medical equipment, position transmitters, media parameters monitoring devices, water preparation and monitoring devices, ship automation facilities, automatic facilities for oil-and-gas producing industry, automatic facilities for power engineering projects, automatic for port terminals, information displaying and processing devices